  • Appliance typeCirculation thermostat
    Class designation acc. DIN 12876III
    Identification according to DIN 12876FL
    Cooling agentR134a
    Cooling agent quantity250 g
    Cooling agent pressure max.20 bar
    Heat output2500 W
    Cooling capacity (@20°C)350 W
    Cooling capacity (@10°C)320 W
    Cooling capacity (@0°C)270 W
    Cooling capacity (@-10°C)190 W
    Cooling capacity (@-20°C)80 W
    Cooling capacity (@ -20°C; 3200 rpm)130 W
    Cooling capacity (@ 20°C; 3200 rpm)400 W
    Working temperature-25 - 200 °C
    Operating temperature min.-25 °C
    Temperature displayyes
    Temperature stability DIN 12876 (@+70°C)±0.01 K
    Temperature controlPID
    Working temperature sensorPT 100
    Safety temperature sensorPT1000
    Working temperature displayTFT
    Safety temperature displayTFT
    Connection for ext. temperature sensorPT 100
    Display resolution0.01 K
    Display for operation with ext. sensoryes
    Warning function optical yes
    Warning function acoustic yes
    Warning function excess temperature yes
    Warning function insufficient temperature yes
    Adjustable safety circuit0 - 260 °C
    Sub-level protectionyes
    Filling volume5 - 7 l
    Pump typePressure- / suction pump
    Pump capacity adjustable yes
    Pump pressure max. (0 liters discharge flow)0.61 bar
    Pressure pump (suction side) (0 liter flow)0.45 bar
    Flow rate max. (0 bar back pressure)31 l/min
    Pump connectionM16x1
    Bath depth min.150 mm
    Height including bath172 mm
    Width bath opening82 mm
    Calibration optionyes
    Technical data complies with the standardDIN 12876
    Multi-Interfaceyes
    Alarm output (potential-free contact) max. 30 V AC/DC
    Alarm output (potential-free contact) max. 1 A
    Solenoid valve output24 VDC
    Solenoid valve output max.0.8 A
    Standby input5 VDC
    Noise level 55 dB(A)
    Dimensions (W x H x D)275 x 690 x 490 mm
    Weight46.5 kg
    Permissible ambient temperature5 - 32 °C
    Permissible relative humidity80 %
    Protection class according to DIN EN 60529IP 21
    RS 232 interfaceyes
    USB interfaceMicro-USB
    Voltage230 V
    Frequency50/60 Hz
    Power input3000 W