HABITAT cell

Scope of delivery

  • HABITAT cell S099
  • Tablet with software for bioreactor
  • HA.pl. Peltier cooling element
GUARANTEED SAFE CHECKOUT
  • Stripe
  • Visa Card
  • MasterCard
  • PayPal
  • Apple Pay
SKU: GBA0010007533 Category:

Description

The HABITAT cell control tower package is an advanced bioreactor system specifically designed for cell culture applications, offering precise monitoring and control of bioprocesses. It includes a control unit with comprehensive connectivity options for gas supply, liquid addition, multi-parameter sensors, and temperature control, along with a user-friendly tablet interface for seamless bioreactor operation.

Control and monitoring
The intuitive software allows easy control and monitoring of the bioprocess and associated parameters via 10.4-inch tablet. Users can select from batch, fed-batch, and perfusion/continuous operating modes, catering to various cell culture techniques. The innovative Chaotic Mixing function enables efficient and homogeneous mixing when required. The control software provides clear calibration management, real-time data visualization, and complete process documentation, ensuring compliance with FDA CFR Part 11 requirements (optional).

Gas Supply
Precise and individually adjustable gassing is achieved through built-in mass flow controllers for four separate gas lines (N2, O2, air, and CO2), enabling flow rates of 0.01 – 2 L/min tailored to cell growth requirements.

Liquid Supply
Four integrated Watson Marlow pumps with adjustable direction and speed offer versatility in pumping in and out various liquids, such as acid, base, anti-foam agents, and feeding solutions for mammalian cell culture media management.

Sensors
HABITAT Bioreactor comes equipped with advanced sensors to precisely monitor and control critical bioprocess parameters. The following sensors are included in our standard sales vessel packages:
• pH
• DO (dissolved oxygen)
• temperature
• filling level
• foam
Optional sensors: CO2, Conductivity, Turbidity, Viable cell mass, Redox, Off-gas analysis and more.

Temperature control
The constant and precise temperature control during the cultivation is guaranteed by the heating sleeve, which is adapted to the respective vessel size. (Heating sleeve is included with the HABITAT cell vessel package).

All features at a glance
• compact, space-saving design, dimensions: 223 x 402 x 450 mm (WxDxH)
• latest technology sensors for better insights and more control
• large screen tablet for clear operation (10,4 inch)
• intuitive, easy-to-use control software that provides simple initial setup and intuitive user assistance throughout the process
• 4 integrated fast-load pumps (Watson Marlow)
• cross platform connectivity: USB, PC, RS232, OPC-UA, ethernet, external signal input, external pump, single-use unit, thermostat
• data storage
• gas supply with 4 built-in mass flow controllers: e.g. N2, O2, air, and CO2
• status LED display: direct error display by indicator light
Note: Only a control tower package and the separately available vessel package make up a functional unit.

Technical Data

Useable volume max.1000 ml
Useable volume with disperser tool min.500 ml
Useable volume without disperser tool min.300 ml
Working temperature max.120 °C
Attainable vacuum25 mbar
Viscosity max.100000 mPas
Speed range10 - 150 rpm
Support rod diameter (with integrated fastening on stand)16 mm
Material in contact with mediumAISI 316L, 1.4571, borosilicate glass 3.3, PTFE, PEEK, FKM
Reactor vessel openings (units/standard)1/NS 29/32, 3/NS 14/23
Permissible ON time100 %
Temperature measurement resolution0.1 K
Nominal torque3 Nm
Heat output1000 W
Cooling medium permissible operating pressure1 bar
Speed min (adjustable)10 rpm
Heating temperature max.180 °C
Adjustable safety circuit47 - 225 °C
Adjustable safety circuit, temperature deviation±10 - ±20 K
Cooling medium temperature min.3 °C
Set temperature resolution±1 K
Heat control accuracy of medium with sensor±0.2 K
Speed deviation±5 rpm
Speed display7 segment LED
Heat controlLED
Connection for ext. temperature sensorPT 100
Timeryes
Timer display7 segment LED
Speed adjustment1 RPM steps
Type of coolingflow
Temperature displayyes
Safety temperature displayLED
Display for operation with ext. sensoryes
Safety cutoutyes
Dimensions (W x H x D)443 x 360 x 295 mm
Weight22 kg
Permissible ambient temperature5 - 40 °C
Permissible relative humidity80 %
Protection class according to DIN EN 60529IP 21
RS 232 interfaceyes
USB interfaceyes
Voltage230 V
Frequency50/60 Hz
Power input1200 W